# Calibrating the K-TCxx

You can calibrate the K-TCxx to Improve the Accuracy of the measurements. The software allows you to cycle through a number of measurement points. Once done, committing the calibration will apply the calibration to the K-TCxx, and measurement accuracy will be improved.

<figure><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2FTTRBEszlndQbSLqe77oU%2Fimage.png?alt=media&#x26;token=a715fe57-085e-4ec4-af62-8a3779be8afb" alt=""><figcaption></figcaption></figure>

<table data-header-hidden><thead><tr><th width="109"></th><th></th></tr></thead><tbody><tr><td><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2FeiV24WLD8B1DveR5nn0j%2Fimage.png?alt=media&#x26;token=7d2ba217-357c-49f6-acda-aedb2c207f26" alt="" data-size="original"></td><td><p>Pressing the Load Button Loads the calibration contained in the K-TCxx and displays them on the screen.</p><p>You may use this function if you wish to check that the K-TCxx is calibrated.</p></td></tr></tbody></table>

Alternatively, press the start button and attach the appropriate thermocouple test signals to the inputs of the K-TCxx in order to cycle through the test points in order to calibrate the K-TCxx for maximum accuracy:

Select the Channel of the Calibration Point you wish to calibrate by either clicking the Radio Button on the channel.

<figure><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2F5TH5NvoK8XmyyI6F0c0M%2Fimage.png?alt=media&#x26;token=2b30aeb2-b4e9-4c2b-81c6-f08ee4fb246a" alt=""><figcaption></figcaption></figure>

<figure><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2F1iYcF51dwRH5YITnDCWm%2Fimage.png?alt=media&#x26;token=150f5ee5-d6d6-438d-9554-2ec2e9dab6f7" alt=""><figcaption></figcaption></figure>

Or by clicking on the point you wish to calibrate in the field to the right

&#x20;

<figure><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2FbWanuCJWQWAiJ2p0TkMf%2Fimage.png?alt=media&#x26;token=cece3b3f-5fca-43ea-b47d-98b1b7d87e42" alt=""><figcaption></figcaption></figure>

Adjust the input temperature from the calibrated test equipment that you are using to match the voltage for the test point:

&#x20;

![](https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2FLfevmOQkw0aKszh3A5GB%2Fimage.png?alt=media\&token=20dbd892-4753-4577-824c-b93c96444f59) Press the Start button:

If the input temperature is stable and correct, the calibration of the channel will be updated as follows:

![](https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2FQSVTdQSRTopxLOXLotdB%2Fimage.png?alt=media\&token=c613fdcf-6772-4307-8a15-fa668cb13cc8)

If it is not, you will receive a warning:

<figure><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2FB0VIE4wsLO4kGduZhPV0%2Fimage.png?alt=media&#x26;token=741121a9-f2b0-4e54-a48b-0f31fab14e43" alt=""><figcaption></figcaption></figure>

Repeat the steps above until you have calibrated all the calibration points.

<table data-header-hidden><thead><tr><th width="128"></th><th></th></tr></thead><tbody><tr><td><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2Foa3iyzqbcT1jOc3e5qyl%2Fimage.png?alt=media&#x26;token=abc9dbf5-9e51-4939-b536-4d15e0cc9b7f" alt=""></td><td><p>Once you have configured the K-TCxx as desired, click Commit to send the configuration to the K-TCxx. If you make a mistake and wish to clear the calibration, you can also use the Clear button.</p><p> </p></td></tr></tbody></table>

This will save the calibration to the K-TCxx and create a report detailing the calibration:

<figure><img src="https://1039220062-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FRhWtYvkHNZy7Jvdnc3wk%2Fuploads%2F1lM1T6gTm7jRwAZ6UcFb%2Fimage.png?alt=media&#x26;token=75b94623-cdd1-4f96-a6bd-7a0e2f3c231f" alt=""><figcaption></figcaption></figure>