Page cover image

K-Box

INF2201

K-Box Multi-DAQ Modules aid in measuring Thermocouples, Voltages, Currents and PWM. These include our unique power down and WakeOnCAN feature and can be stacked with all of our K-series instrumentation modules.

K-Box transmits the Data periodically on the CAN Bus for accurate sensor measurement.

541KB
KBox Flyer.pdf
pdf

Key Features

  • Software switchable voltage input ranges from ±80V to ±10V

  • Up to 8 thermocouple connections at up to 10 Hz sampling rate

  • 8 Analog inputs with variable input sampling rates (8 channels at 1kHz, 4 channels at 2kHz and 2 Channels at 10kHz)

  • Very high accuracy-Analog ±0.0015%,Thermocouples ±1°C

  • PWM: 3 inputs frequency measurements, counters or pulse measurements

  • Outputs: 4 Relay outputs (optional)

  • Regulated +5V and +24V output power supply for external sensors

  • Supplied with configuration software, Influx K-Cal for Windows® and configurable via a DBC file

  • Instrumentation data time synchronised with recorded vehicle network data via CAN

  • Galvanic isolation between modules (enclosure, power, CAN BUS and Analog input module and thermocouple input module)

  • Measurement accuracy: ±1° C

  • Measurement resolution: 0.1°C

  • Analog channel over-voltage protection ±150 Volt

Advantages

  • Combined Voltage, Thermocouple and PWM/Digital measurements

  • Cold junction compensation for accurate thermocouple measurement

  • High input impedance on Analog inputs

  • WakeOnCAN and power down deep sleep mode

Technical Specification

Technical DataDescription

Power supply

6V to 36V DC.

Interfaces

CAN Bus

PC interfaces

None

Enclosure

Dimension (L115xH79xW105) / Weight 450g / IP65 / ABS

Environmental

-40°C to +85°C Humidity max 90%

Output voltages

5V and 24V (Max current drawn 80 mA combined)

Analog Inputs

Number of channels

8 Bipolar differential inputs

Accuracy

±0.0015%

Software switchable range

±80V, ±40V, ±20V, ±10V

Resolution (ADC)

16 Bit

Max Sampling Rate

1 KHz (all 8 channels), 2 KHz (4Channels), 10 KHz (2 Channels)

Input impedance

> 4 M Ohm

Max input voltage

±75 Analog Ground, ±34 V Analog inputs

Thermocouple Inputs

Number of channels

8 J/K/T-type inputs

Accuracy

±1°C accuracy

Measurement Range

Measurement: -200 °C to 1250 °C

Max Sampling Rate

10 Hz (all 8 channels)

Maximum input voltage

±3.3 V

Digital Input / Output

Number of channels

×4 unipolar single-ended hardware configured as inputs or outputs

Input switching thresholds

Low < 1.5V

High > 2.0V (up to 12V)

Input leakage current

< 10nA

Output states

(Optional) Open collector - 510 Ohm

Output drive capability (OK):

Collector-emitter voltage

45V max

Collector current (DC)

10mA max

Saturation voltage (OK on)

< 0.15V

Equivalent on-resistance

< 510Ohm

Leakage current at OK off

< 5uA

Min-max applied voltage

Digital input -8V to +12V; Digital output 0V to +40V power supply, which limits the current to 10mA

PWM

3 digital input frequency measurements up to 100kHz or pulse measurements. (min 100 Nano sec) (between pulses 10 microseconds)

Detailed Documentation