Calibration

You can calibrate the K-TCxx to Improve the Accuracy of the measurements. The software allows you to cycle through a number of measurement points; once done, the calibration will apply the calibration to the K-TCxx, and measurement accuracy will be improved.

Pressing the Load Button Loads the calibration in the K-TCxx and displays them on the screen.

You may use this function if you wish to check that the K-TCxx is calibrated

Nominal Value

Allows the user to select the value being fed to the thermocouple channel.

Raw Temperature

Displays the RAW temperature data.

Noise

Displays the noise level in the signal.

Settings

Allows the user to set the Nominal temperature values.

Noise Level

Allows the user to set the tolerance level.

Clear Calibration

Clears the existing calibration in the device.

Clear Channel

Clears the calibration for the particular channel.

Alternatively, press the start button and attach the appropriate thermocouple test signals to the inputs of the K-TCxx to cycle through the test points to calibrate the K-TCxx for maximum accuracy.